Technical Information
TI365O/98/en/01.10
185414 01/10 03
Ultrasonic Level Sensor
LUC-M**
Compact transmitters for non-contact
level measurement of fluids, pastes and
coarse bulk materials
Application
Continuous, non-contact level
measurement in fluids, pastes,
sullages and coarse bulk materials
Flow measurement in open channels
and measuring weirs
System integration via:
- HART (standard), 4 mA ... 20mA
- PROFIBUS PA
Maximum measuring range:
LUC-M10:
- 5 m (16.4 ft) in fluids
- 2 m (6.6 ft) in bulk materials
LUC-M20:
- 8 m (26.2 ft) in fluids
- 3,5 m (11.5 ft) in bulk materials
LUC-M30:
- 15 m (49.2 ft) in fluids
- 7 m (23 ft) in bulk materials
LUC-M40:
- 10 m (32.8 ft) in fluids
- 5 m (16.4 ft) in bulk materials
Features
Quick and simple commissioning via
menu-guided on-site operation with
four-line plain text display
Envelope curves on the on-site display
for simple diagnosis
Suitable for explosion hazardous
areas (Gas-Ex, Dust-Ex)
Linearisation function (up to 32 points)
for conversion of the measured value
into any unit of length, volume or flow
rate
Non-contact measurement method
minimizes service requirements
Optional remote display and operation
(up to 20 m from transmitter)
Installation possible from thread G1½
or 1½ NPT upwards
Integrated temperature sensor for
automatic correction of the
temperature dependent sound
velocity
LUC-M10
LUC-M40
LUC-M30
LUC-M20
Seitenansicht 0
1 2 3 4 5 6 ... 33 34

Inhaltsverzeichnis

Seite 1 - Ultrasonic Level Sensor

Technical InformationTI365O/98/en/01.10185414 01/10 03Ultrasonic Level SensorLUC-M**Compact transmitters for non-contactlevel measurement of fluids, p

Seite 2 - Table of contents

Ultrasonic Level Sensor LUC-M**Performance characteristicsDOCT-0805F 01/2010 18541410Performance characteristicsReaction time The reaction time depend

Seite 3 - Function and system design

Ultrasonic Level Sensor LUC-M**Installation conditionsDOCT-0805F 01/2010 18541411Installation conditionsInstallation variantsLUC-M10, LUC-M20For insta

Seite 4

Ultrasonic Level Sensor LUC-M**Installation conditionsDOCT-0805F 01/2010 18541412Installation conditions for level measurements• Do not install the se

Seite 5 - Range in m

Ultrasonic Level Sensor LUC-M**Installation conditionsDOCT-0805F 01/2010 18541413Installation conditions for flow measurements• Install the LUC-M** at

Seite 6

Ultrasonic Level Sensor LUC-M**Installation conditionsDOCT-0805F 01/2010 18541414Blocking distance,nozzle installationInstall the LUC-M** at a height

Seite 7 - Auxiliary energy

Ultrasonic Level Sensor LUC-M**Ambient conditionsDOCT-0805F 01/2010 18541415Ambient conditionsAmbient temperature -40 °C ... +80 °C (233 K ... 353 K)T

Seite 8 - PROFIBUS PA

Ultrasonic Level Sensor LUC-M**Mechanical constructionDOCT-0805F 01/2010 18541416Mechanical constructionDesign, dimensions LUC-M10, LUC-M20Dimensions

Seite 9

Ultrasonic Level Sensor LUC-M**Mechanical constructionDOCT-0805F 01/2010 18541417LUC-M40 with mounting bracketDimensions in mm (inch)LUC-M30Dimensions

Seite 10 - Performance characteristics

Ultrasonic Level Sensor LUC-M**Mechanical constructionDOCT-0805F 01/2010 18541418Flanges for LUC-M40Dimensions in mm (inch)WeightSuitable for A B C D

Seite 11 - Installation conditions

Ultrasonic Level Sensor LUC-M**Mechanical constructionDOCT-0805F 01/2010 18541419Housing design Types of housings• F12 housing with sealed terminal co

Seite 12 - DOCT-0805F 01/2010 185414

Ultrasonic Level Sensor LUC-M**Table of contentsDOCT-0805F 01/2010 1854142Function and system design. . . . . . . . . . . . . 3Measuring principle . .

Seite 13

Ultrasonic Level Sensor LUC-M**Human interfaceDOCT-0805F 01/2010 18541420Human interfaceDisplay and operating elementsThe LCD module LUC-Z15 for displ

Seite 14 - LUC-M10/20

Ultrasonic Level Sensor LUC-M**Human interfaceDOCT-0805F 01/2010 18541421On-site operation Operation with LUC-Z15The LC-Display LUC-Z15 allows configu

Seite 15 - Process conditions

Ultrasonic Level Sensor LUC-M**Certificates and approvalsDOCT-0805F 01/2010 18541422Remote control Operation with PACTwareTM (for communication varian

Seite 16 - LUC-M10 LUC-M20

Ultrasonic Level Sensor LUC-M**Ordering informationDOCT-0805F 01/2010 18541423Ordering informationProduct structureLUC-M10Product structureLUC-M20LUC–

Seite 17 - Mechanical construction

Ultrasonic Level Sensor LUC-M**Ordering informationDOCT-0805F 01/2010 18541424Product structure LUC-M30Product structureLUC-M40LUC–M30– – –Certificate

Seite 18 - 20 (0.79)

Ultrasonic Level Sensor LUC-M**AccessoriesDOCT-0805F 01/2010 18541425Scope of delivery • instrument according to the version ordered• operating manual

Seite 19

Ultrasonic Level Sensor LUC-M**AccessoriesDOCT-0805F 01/2010 18541426Cantilever• The 50 mm (2 in) or 62 mm (2.4 in) orifices serve for the mounting of

Seite 20 - Snap-fit

Ultrasonic Level Sensor LUC-M**AccessoriesDOCT-0805F 01/2010 18541427Wall bracketAdapter flange forLUC-M10/LUC-M20Version with metrical threadVersion

Seite 21 - Envelope

Ultrasonic Level Sensor LUC-M**AccessoriesDOCT-0805F 01/2010 18541428Universal slip-on flange for LUC-M30Remote displayTechnical dataProzess connectio

Seite 22 - Certificates and approvals

Ultrasonic Level Sensor LUC-M**AccessoriesDOCT-0805F 01/2010 18541429Display and operating module LUC-Z15LC display for on-site operation of LUC-M** (

Seite 23 - Ordering information

Ultrasonic Level Sensor LUC-M**Function and system designDOCT-0805F 01/2010 1854143Function and system designMeasuring principleE: empty distance; F:

Seite 24

Ultrasonic Level Sensor LUC-M**Supplementary documentationDOCT-0805F 01/2010 18541430Supplementary documentationOperating manual Depending on the comm

Seite 25 - Accessories

Ultrasonic Level Sensor LUC-M**Supplementary documentationDOCT-0805F 01/2010 18541431Safety Instructions The following safety instructions are supplie

Seite 26

Ultrasonic Level Sensor LUC-M**Supplementary documentationDOCT-0805F 01/2010 18541432Control drawings/installation drawingsThe following control or in

Seite 27

With regard to the supply of products, the current issue of the following document is applicable: The General Terms of Delivery for Products and Servi

Seite 28

Subject to modificationsCopyright PEPPERL+FUCHS • Printed in Germanywww.pepperl-fuchs.comWorldwide HeadquartersPepperl+Fuchs GmbH68307 Mannheim · Ge

Seite 29

Ultrasonic Level Sensor LUC-M**Function and system designDOCT-0805F 01/2010 1854144Equipment architecture 4 mA … 20 mA output with HART protocolThe co

Seite 30 - Supplementary documentation

Ultrasonic Level Sensor LUC-M**InputDOCT-0805F 01/2010 1854145InputMeasured variable The distance D between the sensor membrane and the product surfac

Seite 31

Ultrasonic Level Sensor LUC-M**OutputDOCT-0805F 01/2010 1854146Example (for LUC-M30)For typical solid applications, a certain amount of dust coverage

Seite 32

Ultrasonic Level Sensor LUC-M**Auxiliary energyDOCT-0805F 01/2010 1854147Auxiliary energyTerminal compartment In the F12 housing, the terminals are lo

Seite 33

Ultrasonic Level Sensor LUC-M**Auxiliary energyDOCT-0805F 01/2010 1854148The digital communication signal is transmitted to the bus via a 2-wire conne

Seite 34 - PROTECTING YOUR PROCESS

Ultrasonic Level Sensor LUC-M**Auxiliary energyDOCT-0805F 01/2010 1854149Cable entry • cable gland: M20 x 1.5 (recommended cable diameter 6 mm ... 10